Skip to Main Content (Press Enter)

Logo CNR
  • ×
  • Home
  • People
  • Outputs
  • Organizations
  • Expertise & Skills

UNI-FIND
Logo CNR

|

UNI-FIND

cnr.it
  • ×
  • Home
  • People
  • Outputs
  • Organizations
  • Expertise & Skills
  1. Outputs

A procedure for correcting self-absorption in calibration free-laser induced breakdown spectroscopy

Academic Article
Publication Date:
2002
abstract:
A model of the self-absorption effect in laser-induced plasma has been developed, with the aim of providing a tool for its automatic correction in the Calibration-Free algorithm recently developed for standardless analysis of materials by LIBS (Laser Induced Breakdown Spectroscopy). As a test of the model, the algorithm for self-absorption correction is applied to three different certified steel NIST samples and to three ternary alloys (Au, Ag, Cu) of known composition. The experimental results show that the self-absorption corrected Calibration-Free method gives reliable results, improving the precision and the accuracy of the CF-LIBS procedure by approximately one order of magnitude.
Iris type:
01.01 Articolo in rivista
Keywords:
LIBS; self absorption; curve of growth
List of contributors:
Tognoni, Elisabetta; Cristoforetti, Gabriele; Salvetti, Azenio; Palleschi, Vincenzo
Authors of the University:
CRISTOFORETTI GABRIELE
PALLESCHI VINCENZO
TOGNONI ELISABETTA
Handle:
https://iris.cnr.it/handle/20.500.14243/38811
Published in:
SPECTROCHIMICA ACTA, PART B: ATOMIC SPECTROSCOPY