X-ray dosimeters were developed by tailoring
novel injecting diamond-like-carbon/Pt/Au contacts on singlecrystal
high-purity diamond films. A dark resistivity of
(5.6 ± 0.1) × 1014 ? · cm and no appreciable presence of
deep traps in the bandgap confirmed a very low defect density in
the diamond films. The dosimeters resulted to be primingless (i.e.,
no need of preactivation) and characterized by high sensitivity
[(58.20 ± 3.26) × 10-3 C · Gy-1 · cm-3] and linear response to
X-ray dose rate, produced by a molybdenum target. Transient
X-ray modulated analysis allowed the determination of fast-trap
influence and the estimation of very fast response times (~10-3 s),
at electric fields >= 3 × 103 V/cm.