Characterization of a plasma source for biomedical applications by electrical, optical, and chemical measurements
Academic Article
Publication Date:
2018
abstract:
This paper reports the characterization of a low-power, atmospheric pressure plasma source developed for and successfully tested in applications of plasmamedicine. The source produces plasma by ionizing a helium flow containing traces of air. The effluent gas is thus enriched in reactive oxygen and nitrogen species as well as helium metastable excited states. Electrical measurements, Schlieren imaging, OES and various chemical analyses and probes enabled usto fully characterize the plasma source and to determine some fundamental parameters for treating living samples such as the power feeding the discharge, the optimal source to target distance, the maximum exposure time, and to identify the major chemical species responsible for the biological action on exposed living matter.
Iris type:
01.01 Articolo in rivista
Keywords:
atmospheric plasma jet; optical emission spectroscopy; plasma medicine; reactive oxygen species; Schlieren
List of contributors:
Martines, Emilio; Zuin, Matteo
Published in: