abstract:
This article describes the completely renovated Thomson scattering (TS) diagnostic employed in the modified Reversed Field eXperiment (RFX-mod) since it restarted operation in 2005. The system measures plasma electron temperature and density profiles along an equatorial diameter, measuring in 84 positions with 7 mm spatial resolution. The custom built Nd:YLF laser produces a burst of 10 pulses at 50 Hz with energy of 3 J, providing ten profile measurements in a plasma discharge of about 300 ms duration. An optical delay system accommodates three scattering volumes in each of the 28 interference filter spectrometers. Avalanche photodiodes detect the Thomson scattering signals and allow them to be recorded by means of waveform digitizers. Electron temperature is obtained using an alternative relative calibration method, based on the use of a supercontinuum light source. Rotational Raman scattering in nitrogen has supplied the absolute calibration for the electron density measurements. During RFX-mod experimental campaigns in 2005, the TS diagnostic has demonstrated its performance, routinely providing reliable high resolution profiles. (c) 2007 American Institute of Physics.
Iris type:
01.01 Articolo in rivista
Keywords:
THOMSON-SCATTERING; REVERSED-FIELD-PINCH; RAMAN-SCATTERING; CORE TRANSPORT; CALIBRATION
List of contributors:
Pasqualotto, Roberto
Authors of the University:
PASQUALOTTO ROBERTO
Handle:
https://iris.cnr.it/handle/20.500.14243/26190
Published in:
REVIEW OF SCIENTIFIC INSTRUMENTS
Journal
  • Use of cookies

Powered by VIVO | Designed by Cineca | 26.5.0.0 | Sorgente dati: PREPROD (Ribaltamento disabilitato)